Procedure in case of malfunctions
User Manual corpuls
3
264 ENG - Version 2.1 P/N 04130.2
Notification in the
message line and
information in the
protocol
Explanation/Measure
Charging
Message in the message line that the defibrillator is being charged.
Wait until the charging process is finished and the defibrillator
signals readiness to shock by issuing the ready-signal.
Charging not possible
A technical error occurred.
The temperature of the charging generator has exceeded a limit
value due to several discharges.
Leave the corpuls
3
to cool down.
If the error persists, contact authorised sales and service partners.
Check modules
If there is no connection between the monitoring unit and the patient
box and/or the defibrillator/pacer at the time of switch-off or if there
is a timing problem between the modules, this is communicated to
the user by a message on the display.
In this case, disconnect the the modules and check if all modules
are switched off. If this is not so, switch off the still running modules
via the respective On/Off key (hold down for 3 seconds).
Check therapy electrodes
The therapy electrodes are not connected correctly to the therapy
master cable.
Check the connection of the plugs and adjust if necessary.
Use replacement corPatch electrodes (see also chapter 5.1.1
Types of Therapy Electrodes, page 63)
If the message persists, contact sales and service partners.
Circuit open - Check pacer
electrodes
The corPatch electrodes are not connected correctly to the
patient or they have a too high resistance to the patient’s skin.
Stimulation is not possible.
Check the expiry date of the electrodes, if they are dried out and if
they are placed correctly on the patient.
Code changed
Confirmation that the access code has been successfully changed.
Code invalid - Retry?
--
Code mismatch - Retry?
The verification of the new access code did not match the earlier
entry.
To repeat the process, confirm the prompt and enter the access
code again.
Configuration loaded
--
Configuration reset
--
Configuration stored
--
Confirm alarms.
Message in the status line that the alarm messages contained in the
following alarm list can be confirmed after the user has acknowledged
them and has taken the resulting measures on the patient.