Procedure in case of malfunctions
ENG - Version 2.1 P/N 04130.2 265
Notification in the
message line and
information in the
protocol
Explanation/Measure
Connect defibrillator and
monitoring unit
Message in the message line of the monitoring unit.
The wireless communication between the defibrillator and the
monitoring unit/patient box has been interrupted or could not be
established:
Make sure that the distance between the modules is not more than
10 m and that no barriers are compromising the wireless
connection. Use corpuls
3
as a compact device if necessary.
Connect ECG cable
--
Connect pacer cable
User prompt to connect the therapy master cable to the therapy
electrodes.
Connect therapy electrodes
Connect the corPatch electrodes or shock paddles to the therapy
master cable of the corpuls
3
.
If the alarm message persists, connect immediately the reserve
corPatch electrodes.
Connected to server
Status message during telemetry.
Connection to the server could be established.
Data transmission is active.
Continuing pacing
User prompt to confirm the continuation of stimulation (pacer
therapy).
D-ECG ([NUMBER]) not
saved ([NUMBER])
The recorded D-ECG could not be saved because
the CF card is possibly full,
the CF card is possibly not inserted correctly
the CF card is possibly faulty or
the writing process onto the CF card has failed.
After deleting or exchanging the CF card, repeat the process. If the
error persists, contact authorised sales and service partners.
D-ECG [NUMBER]) saved
Event in the protocol that a D-ECG has been saved.
The number of the D-ECG is shown in brackets.
D-ECG fax sent
Confirmation that the fax has been successfully sent.
D-ECG printout
Event in the protocol, recording that a D-ECG has been printed out.
D-ECG sent to server
Event in the protocol, recording that a D-ECG has been successfully
sent to the server.
Defibrillation only possible
in manual mode
User prompt to switch from AED mode to manual defibrillation
mode.
The therapy electrodes used are not authorised for use in AED
mode.
Defibrillator activated (AED)
Event in the protocol that the defibrillator has been activated in AED
mode.
Defibrillator activated
(MAN)
Event in the protocol that the defibrillator has been activated in
manual mode.
Defibrillator deactivated
Event in the protocol, recording the leaving the defibrillator mode.