Procedure in case of malfunctions
User Manual corpuls
3
266 ENG - Version 2.1 P/N 04130.2
Notification in the
message line and
information in the
protocol
Explanation/Measure
Defibrillator powered on
Event in the protocol, recording the switching on of the
defibrillator/pacer.
Demo mode off
The DEMO mode has been switched off.
Demo mode on
The DEMO mode has been switched on.
Check if curves and parameters are displayed.
Disconnect defibrillator and
monitoring unit
Message in the message line of the monitoring unit.
Communication in the connected state between the defibrillator and
monitoring unit/patient box is cut off or could not be established:
Check if one of the two infrared interfaces is covered or dirty.
Do not touch or move
patient
The ECG analysis is being performed.
Do not touch or move patient.
Follow the instructions on the display.
ECG analysis result;
>>> Shock not
recommended
ECG analysis result;
>>> Shock recommended
Event in the protocol, recording the result of the ECG analysis.
ECG analysis started
Event in the protocol, recording that an ECG analysis has been
performed in AED mode.
Energy selection
[NUMBER] J
Event in the protocol, recording the energy selected.
Enter code: [NUMBER]
[NUMBER] [NUMBER]
[NUMBER]
User prompt to enter the access code.
Enter new code:
[NUMBER] [NUMBER]
[NUMBER] [NUMBER]
User prompt to enter a new access code.
Error in device (BIM)
The corpuls
3
is not functioning correctly and must not be used.
Contact authorised sales and service partners.
Error in device (DEFI)
The corpuls
3
is not functioning correctly and must not be used.
Contact authorised sales and service partners.
Error in device (MAN-BIM)
The corpuls
3
is not functioning correctly and must not be used.
Contact authorised sales and service partners.
Error in device (PIF)
The corpuls
3
is not functioning correctly and must not be used.
Contact authorised sales and service partners.
Event recorded
--
Failure HES data
transmission
An error occurred during data transmission of the ECG measuring
and interpretation software HES
®
.
Data transmission was not successful.
Repeat procedure
If the data transmission fails repeatedly, contact authorised sales
and service partners.