User Manual corpuls
3
Procedure in case of malfunctions
ENG - Version 2.1 – P/N 04130.2 267
message line and
information in the
protocol
Fax connection not possible
Possibly a wrong fax number has been dialled.
Repeat dialing.
Status indication during fax transmission
[MODULE] - Pairing
impossible
The connection authorisation (Pairing) has failed.
Due to different hardware versions the modules cannot be
connected.
Contact authorised sales and service partners.
The electrical resistance of the patient (impedance) is to high for the
selected settings. An intensity for stimulation has been selected that
cannot be reached with the impedance present.
Make sure that the corPatch electrodes are tightly and completely
attached to the patient's skin.
In case of excessive hair on the patient, shave the required area on
the skin. If necessary, use new corPatch electrodes.
To perform a pacer therapy adaequate for the patient, a higher
stimulation intensity must be selected.
Perform medical measures as needed.
Keyboard lock has been engaged.
key HOME pressed to
unlock
--
Keyboard lock has been disengaged.
User prompt to confirm the locking of the keyboard.
Metronome child [TEXT]
Event in the protocol, recording the switching on, switching off and
the mode of the metronome.
Event in the protocol indicating the ending time of the device mission.
Event in the protocol indicating the starting time of the device mission.
Event in the protocol, recording the switching on of the monitoring
unit.
[Last name], [name]
An insurance card has been recognised.
New code invalid - Retry?
User prompt to re-enter the new access code, if the repetition was
wrong.
[NUMBER] / [NUMBER]
([NUMBER]) mmHg
Result of the NIBP measurement in the protocol.