Procedure in case of malfunctions
User Manual corpuls
3
268 ENG - Version 2.1 – P/N 04130.2
message line and
information in the
protocol
defibrillator unit
Message in the message line of the monitoring unit.
The wireless communication between the defibrillator and the
monitoring unit/patient box has been interrupted or could not be
established:
Make sure that the distance between the modules is not more than
10 m and that no barriers are compromising the wireless
connection. Use corpuls
3
as a compact device if necessary.
Communication in the connected state between the defibrillator and
monitoring unit/patient box is cut off or could not be established:
Check if one of the two infrared interfaces is covered or dirty.
Message in the message line of the monitoring unit.
The wireless communication between the patient box and the
monitoring unit/defibrillator has been interrupted or could not be
established:
Make sure that the distance between the modules is not more than
10 m and that no barriers are compromising the wireless
connection. Use corpuls
3
as a compact device if necessary.
Communication in the connected state between the defibrillator and
monitoring unit/patient box is cut off or could not be established:
Check if one of the two infrared interfaces is covered or dirty.
mode
For operating the pacer in DEMAND mode the 4-pole ECG monitoring
cable has to be connected to the patient and to the corpuls
3
.
The pacer is stimulating.
available
The pacer of the corpuls
3
is activated and the selected intensity is
more than 0 mA.
A D-ECG cannot be performed under these circumstances.
An error occurred during pacer therapy.
The pacer is switched off. Stimulation is interrupted.
Treat patient and take measures as needed.
Contact authorised sales and service partners.
[NUMBER]/min selected
Event in the protocol, recording the pacer frequency selected.
mA selected
Event in the protocol, recording the pacer intensity selected.
selected
Event in the protocol, recording the switch between the pacer modes
FIX and DEMAND. (No alarm!)
Confirmation that the pacer has been switched off and is not
stimulating.
[NUMBER] mA, mode
[NUMBER]
Status indication of a performed pacer therapy in the protocol.
The stimulation frequency, intensity and the selected pacer mode
are indicated.
Event in the protocol, recording how long the pacer therapy has been
interrupted.
Message in the message line that the pacer is running.