User Manual corpuls
3
Appendix
ENG - Version 2.1 – P/N 04130.2 311
I Guidelines and Manufacturer’s Declaration
Electromagnetic emission
The corpuls
3
is intended for operation in the electromagnetic environment indicated below. The
operator or the user has to make sure that the corpuls
3
is used in such an environment.
Emission measurements Compliance Electromagnetic environment -
guidelines
HF emissions according to
CISPR 11
Group 1 The device uses HF energy only for
its internal function. Consequently, its
HF emission is very low and it is
unlikely to interfere with neighbouring
electronic devices.
HF emissions according to
CISPR 11
Class B
The corpuls
3
is suitable for use in
all facilities, including living areas and
those which are directly connected to
the public mains supply.
Furthermore, it is suitable for use in
vehicles, aeroplanes and on ships.
Emission of harmonic oscillations
according to
IEC 61000-3-2
Not applicable
Voltage fluctuations/flicker
according to IEC 61000-3-3
Not applicable
-38 Electromagnetic emission
Electromagnetic interference immunity
The corpuls
is intended for operation in the electromagnetic environment indicated below. The
operator or the user has to make sure that the corpuls
3
is used in such an environment.
Interference
immunity tests
IEC 60601 test
level
Compliance
level
Electromagnetic environment -
guidelines
discharge (ESD)
according to IEC
61000-4-2
contact discharge
± 8 kV
aerial discharge
contact discharge
± 8 kV
aerial discharge
Floors should be made of wood,
concrete or metal or be covered
with ceramic tiles. If the floor is
covered with synthetic material,
the relative humidity should be at
least 30%.
electrical
interference/bursts
according to IEC
61000-4-4
leads
± 1 kV for
connecting leads
connection
Not applicable
The quality of the power supply
should correspond to that of a
typical business or hospital
environment.
IEC 61000-4-5
voltage
± 2 kV common
mode voltage
mode voltage
± 2 kV common
mode voltage
The quality of the power supply
should correspond to that of a
typical business or hospital
environment.