Procedure in case of malfunctions
User Manual corpuls
3
262 ENG - Version 2.1 – P/N 04130.2
Malfunction Possible cause Measure
The printout contains 6
ECG curves of the DE-
recording with a
dashed line or dashed
test signal
The booting process of the
corpuls
3
is not yet
finished.
Wait until the corpuls
3
has booted
completely.
The printer dispenses
paper in an
uncontrolled way.
The corpuls
3
falsely
assumes a paper jam.
Press the Print key again.
If the paper dispensal cannot be stopped,
open the printer flap.
The printer flap has
come loose from its
lock
Excessive use of strength
when opening the printer
flap.
Check if there is visible damage to the
printer flap, printer roll or the hinges.
If there is no visible damage, position the
corpuls
3
on its side and let the printer flap
click into place in its guide rails carefully.
If the error persists, contact sales and
service partners.
The corpuls
3
has been
dropped with the printer
flap open.
10-12 Printer malfunctions
Malfunction Possible cause Measure
corpuls
3
cannot be
switched on.
No power supply is present. Connect the mains charger.
For operation as compact device insert at
least one charged battery.
The patient box or
defibrillator is already
switched on.
Separate corpuls
3
into its three modules
and check whether the patient box or the
defibrillator are switched on (see Status
LEDs, p. 32)Status LEDsStatus LEDs.
If this is not the case, switch on the
remaining devices individually.
The defibrillator/ pacer,
monitoring unit, patient
box cannot be
switched on.
There is no connection to
the power supply.
Connect the mains charger.
No batteries are inserted,
or the batteries are flat.
For operation as compact device insert at
least one charged battery.
Charging is not
possible in spite of
mechanical connection
of the magnetic clip.
A foreign body is present
on the magnetic contact
surface (e.g. a paperclip).
Remove the foreign bodies from the contact
surfaces.
The battery discharges
very rapidly.
The battery shows signs of
wear.
Replace the battery if necessary.
10-13 Malfunctions in energy management