Procedure in case of malfunctions
ENG - Version 2.1 – P/N 04130.2 261
Malfunction Possible cause Measure
The pressure channel is not
calibrated.
Calibrate the pressure channel.
10-11 Malfunctions during IBP monitoring
Malfunction Possible cause Measure
No printout follows in
spite of pressing the
printer key.
The roll of paper in the
printer is used up.
Insert a new roll.
The paper has jammed
during the printing process.
Open the paper flap and remove the paper
jam.
To do this, pull the paper forwards, close
the paper flap and tear the paper off on the
cutting edge.
The paper has become
wound around the printer
roller.
Open the paper flap.
Pull the paper wound around the paper
roller carefully off the roller with both hands.
Close the paper flap and tear off the paper
on the cutting edge.
The paper is not correctly
inserted.
Insert the paper correctly (see chapter 9.5,
Fig. 9-2, p. 217).
The printout is of poor
quality.
The printing head is
possibly dirty.
Clean the printing head carefully with an
alcohol-impregnated cloth.
The printer flap is not fully
locked.
Firmly lock the printer flap on both sides
(see chapter 9.5, p. 217).
No event ECG is
available on printing
the protocol.
No CompactFlash
®
card is
inserted.
Fully insert the CompactFlash
®
card.
Only use original corpuls
3
-
CompactFlash
®
cards.
The CompactFlash
®
card is
full.
Save the data on another data carrier and
then delete the data on the CompactFlash
®
card and reinsert it.
The CompactFlash
®
card
cannot be read.
The CompactFlash
®
card is not formatted
correctly or no original corpuls
3
CompactFlash
®
card was used. Use a
correctly formatted original corpuls
3
CompactFlash
®
card.
No original corpuls
3
CompactFlash
®
card was
used.
It is essential to use the original corpuls
3
CompactFlash
®
cards for safety reasons.
When printing out the
protocol, instead of
parameter values
there are question
marks.
There are no minute mean
values yet.
Print the protocol at a later point of time, if
possible.