Intended Use
ENG - Version 2.1 P/N 04130.2 5
If monitoring functions are performed, the patient’s condition must also be
regularly monitored even when the alarm function is enabled.
The corpuls
3
is not intended for
operation in the vicinity of readily inflammable substances,
setup and operation under the influence of strong electromagnetic fields,
which occur in the direct vicinity of radio masts, switched-on nuclear
magnetic resonance tomography units, high voltage installations and
overhead power lines,
operation in the vicinity of therapeutic radiation units (e.g. tumor
treatment),
operation in connection with a high frequency surgical device,
operation in a monoplace hyperbaric chamber (option HBO),
operation in a multiplace hyperbaric chamber with more than 3 barg
and/or more than 23 % oxygen concentration (option HBO).
The pacer function must not be used near high frequency surgical devices or
microwave therapy devices.
Individual modules must not be used without batteries inserted.
Defibrillation and cardioversion must not be performed without protective
measures (see chapter 5.3.1 Information on the AED Mode, p. 68 and 5.4.1
Information on Manual Defibrillation and Cardioversion, p. 74):
on a metal surface;
on a wet surface.
The defibrillator must only be used for defibrillation and cardioversion and must
not be used as a stimulation current device or as a pacer.
The pacer may only be used as a transcutaneous pacer.
The pacer must not be used as an intracardial defibrillator.
The corpuls
3
may not be used simultaneously on two or more patients.
The manufacturer cannot accept any liability for damage occurring as a result of
failure to use corpuls
3
as intended.
Usage other than
as intended