User Manual corpuls
3
Maintenance and tests
ENG - Version 2.1 – P/N 04130.2 209
Functional
test
Description Measures by the user Correct result
Function
check of the
defibrillator/
pacer
When using shock paddles:
Insert shock paddles into
the shock paddle holder.
The shock paddles must be
heard to click into place.
Connect the shock paddle
cable to the therapy master
cable.
Select the manual mode of
the defibrillator.
Select an energy of 200 J.
Load energy.
Release shock.
Also check reserve shock
paddles, if necessary.
By pressing the shock paddle
buttons a confirmative tone
sounds.
Energy is internally discharged
via the test contacts.
No error message is issued.
When using corPatch
electrodes:
Insert the plug of the
therapy master cable into
the test contact socket at
the cable base.
Select the manual mode of
the defibrillator.
Select an energy of 200 J.
Load energy.
Release shock.
Energy is internally discharged
via the test contact socket.
No error message is issued.
Function
check of the
defibrillator/
pacer SLIM
For the function check the
testbox (P/N 04310) is
needed:
Make sure that a 1.5 V
battery is inserted in the
testbox.
Connect testbox to the
therapy socket.
Switch on testbox.
Select the manual mode of
the defibrillator.
Select an energy of 200 J.
Load energy.
Release shock.
After performing the
function check, switch off
testbox and remove it from
the defibrillator/pacer SLIM.
Energy is internally discharged
via the test box.
No error message is issued.