Operation Monitoring and Diagnosis
User Manual corpuls
3
116 ENG - Version 2.1 P/N 04130.2
Warning
The oximeter may not be operated in the vicinity of ionising (radioactive)
radiation, because the measured values might be falsified.
Caution
Refer to the list of approved accessories (see chapter 9.8 Approved
Accessories, Spare Parts and Consumables, p. 224) for approved sensors
from the company Masimo. Sensors from other manufacturers are not
supported by the device and are not permissible.
The license agreement allows an extension of the available oximetry measuring
options which can be activated by a service technician.
Older Masimo SET oximetry sensors can only be used in connection with an
adapter cable (see chapter 9.8 Approved Accessories, Spare Parts and
Consumables, p. 224).
Simultaneous measurement of SpCO and SpHb is not possible, due to different
sensor types.
A low confidence of the measured value is indicated by a question mark instead
of the bell symbol in the upper right corner of the parameter field.
After connecting the oximetry sensor the oximeter calibrates automatically. This
procedure may take up to 120 seconds for the SpO
2
measurement, during
which an hourglass symbol is displayed in the upper right corner of the
parameter field.
It is recommended to connect SpHb sensors to the corpuls
3
only if needed, to
avoid premature expiration of the sensor.
See the manufacturer’s operating manual for further information on the sensors.
These manuals must be read carefully before use.
The oximeter is covered under one or more of the following U.S.A. patents:
5,758,644, 6,011,986, 6,699,194, 7,215,986, 7,254,433, 7,530,955 and other
applicable patents listed at: www.masimo.com/patents.htm
Purchase or possession of this Masimo Rainbow SET® Oximeter confers no
express or implied licence to use the device with any non-approved sensors or
cables that would fall, alone or in combination with this device, under any patent
in reference to this device.