EasyManua.ls Logo

GS corpuls3 - Page 142

GS corpuls3
346 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation Monitoring and Diagnosis
User Manual corpuls
3
130 ENG - Version 2.1 P/N 04130.2
Caution
Before starting the NIBP measurement, make sure that the selected initial
pressure is appropriate for the patient.
1. In monitoring mode press the softkey [NIBP].
2. Press the softkey [Mode] repeatedly until the required mode "Adult“, "Child"
or "Neonate" appears.
3. To start an individual measurement, press the softkey [Start].
4. The NIBP cuff is inflated and measurement is performed automatically.
5. A measurement in progress can be interrupted with the softkey [Stop].
Adjust the initial pressure of the monitoring at approx. 30 mmHg above the
expected systolic value or use the default settings.
Immediately after the measurement the softkeys are greyed out. Another
measurement is possible only after 5 seconds.
To additionally display the measured value in a parameter field exit the NIBP
user surface with the softkey [Close].
1. Select a parameter field for display of the non-invasive blood pressure and
open the parameter context menu.
2. Assign NIBP monitoring to the selected parameter field.
3. Alternatively, individual parameter fields can be configured to display the
systolic (NIBP SYS), diastolic (NIBP DIA) and the mean arterial (NIBP
MAD) pressure values.
Fig. 6-31 NIBP monitoring parameter fields
1 Systolic value
2 Diastolic value
3 Symbol for switched-on alarms
4 Mean arterial pressure
5 Upper alarm limit
6 Lower alarm limit
Note
Note

Table of Contents

Other manuals for GS corpuls3