EasyManua.ls Logo

GS corpuls3 - Page 296

GS corpuls3
346 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix
User Manual corpuls
3
284 ENG - Version 2.1 P/N 04130.2
Pre-set Alarm Limits
Measurement value
HR 1/min 50 120
SpO
2
% 90 -
PP 1/min 50 120
SpCO % - 10
SpHb g/dl 10 17
SpHb mmol/l 6.2 10.6
SpMet % - 3
CO
2
mmHg 30 50
RR 1/min 8 18
NIBP mmHg SYS 80
DIA 40
SYS 200
DIA 100
P1 mmHg SYS 80
DIA 50
SYS 180
DIA 100
P3 mmHg SYS 80
DIA 50
SYS 180
DIA 100
P2 mmHg SYS 80
DIA 50
SYS 180
DIA 100
P4 mmHg SYS 80
DIA 50
SYS 180
DIA 100
T1 °C 34.0 39
T2 °C 34.0 39
Table A
-4 Pre-set alarm limits