ENG - Version 2.1 P/N 04130.2 313
Electromagnetic interference immunity
Radiated HF
interference
according to
IEC 61000-4-3
10 V/m
80 MHz to 2.5 GHz
3 V/m
ECG, oximetry monitor:
P
d
0
.4=
for 80 MHz to 800 MHz
Pd 7.7=
for 800 MHz to 2.5 GHz
With magnetic forces of > 3 V/m,
disturbances in the ECG signal
may selectively occur.
10 V/m
Defibrillator/Pacer: no
unintentional change in condition
Pd 2.1=
for 80 MHz to 800 MHz
Pd 3.2=
for 800 MHz to 2.5 GHz
20 V/m
Defibrillator:
no unintentional energy release
Pd 6.0=
for 80 MHz to 800 MHz
P
d 2.1=
for 800 MHz to 2.5 GHz
P being the maximum nominal
output of the transmitter in watts
(W) according to the transmitter
manufacturer’s indications and d
being the recommended
protection distance in metres
(m).
b
The magnetic force of stationary
radio transmitters should be lower
than the level of compliance
d
for
all frequencies according to an
on-site test
c
Interference is possible in the
vicinity of devices that bear the
following pictorial symbol:
Fig. 10-1 Radio transmitter