User Manual corpuls
3
Appendix
ENG - Version 2.1 – P/N 04130.2 313
Electromagnetic interference immunity
Radiated HF
interference
according to
IEC 61000-4-3
10 V/m
80 MHz to 2.5 GHz
3 V/m
for 80 MHz to 800 MHz
for 800 MHz to 2.5 GHz
With magnetic forces of > 3 V/m,
disturbances in the ECG signal
may selectively occur.
10 V/m
unintentional change in condition
for 80 MHz to 800 MHz
for 800 MHz to 2.5 GHz
20 V/m
no unintentional energy release
for 80 MHz to 800 MHz
for 800 MHz to 2.5 GHz
P being the maximum nominal
output of the transmitter in watts
(W) according to the transmitter
manufacturer’s indications and d
being the recommended
protection distance in metres
(m).
b
The magnetic force of stationary
radio transmitters should be lower
than the level of compliance
d
for
all frequencies according to an
on-site test
c
Interference is possible in the
vicinity of devices that bear the
following pictorial symbol:
Fig. 10-1 Radio transmitter