Operation Therapy
User Manual corpuls
3
64 ENG - Version 2.1 P/N 04130.2
The internal shock spoons consist of the shock spoon-electrodes and the
handles. Before use the electrodes have to be screwed onto the handles. The
handles are already connected to an electrode cable and only need to be
connected to the therapy master cable of the defibrillator/pacer.
WARNING
To guarantee a defibrillation protection for patients, users and third parties,
use exclusively the accessories indicated in the list of authorised accessories
(see chapter 9.8 Approved Accessories, Spare Parts and Consumables,
page 224).
WARNING
The following safety instructions which can also be found on every pouch of
the corPatch electrodes have to be complied with when using the
corPatch electrodes:
Do not crush, bend, or fold electrodes or store them under heavy
objects.
Do not open pouch until ready for use.
Do not use corPatch electrodes if gel is dry.
Do not use additional gel on corPatch electrodes.
Do not overlap corPatch electrodes.
Use separate ECG electrodes when performing non-invasive pacing.
Do not touch patient during defibrillation.
Do not discharge shock paddles through corPatch electrodes.
Keep corPatch electrodes clear of any other electrodes or metal
parts in contact with the patient.
Not following these instructions for the corPatch electrodes or any other
misuse or misapplication of the corPatch electrodes may result in severe
patient burns or ineffective therapy
Caution
Do not use corPatch electrodes if
the packaging is damaged or opened;
the expiry date indicated on the packaging has passed;
the electrode or the connecting lug is bent.
Caution
Replace corPatch electrodes for adults at the latest after:
24 hours or 50 shocks;
8 hours of continuous pacer operation.
Caution
When placing the corPatch electrodes on the patient’s skin, make sure that
no air pockets are included inside the adhesive surface. Apply corPatch
electrodes from the centre outwards.
WARNING
Before using the shock spoons, please read and understand the additional
user manual (P/N 04137.02).
Internal shock
spoons