Remote Control of the Instrument through the Internet Browser (LAN Communications Only)
If no HTTP screen is displayed
(1) Verify the settings of the Internet browser.
1
On the Internet Explorer settings, click
Tools-Internet Options.
2
On the Advanced tab, enable the “Use
HTTP1.1” and disable the “Use HTTP1.1
through proxy connections.”
3
Under LAN settings on the Connections
tab, disable the Proxy server settings.
(2) Verify the LAN settings.
1
Check the instrument’s LAN settings
and the computers IP address.
See “Settings (SET UP Screen)” (p. 138).
2
Verify that the LINK LED on the LAN
interface is on and that the WEB mark is
being shown on the instrument’s screen.
See “Connection” (p. 140) and “1.6 On screen
indicators” (p. 35).
3
Under LAN settings on the Connections
tab, disable the Proxy server settings.
IMPORTANT
After conguring the LAN settings, be sure to turn the instrument off and on. If this is not
performed, the changed LAN setting will not be enabled and communication will not be possible.
Remote operation
1
Click
2
(If the password has already been set)
2. Click
1. Enter the password
If no password has been set or the password
has been set to “0000” (four zeros), this screen
will not be displayed. The default password is
“0000.”
The same screen and control panel being
displayed on the instrument will be displayed on
the browser.
12