193
15
Maintenance and Service
WARNING
Customers are not allowed to modify, disassemble, or repair the instrument. Doing
so may cause re, electric shock, or injury.
Calibrations
The calibration frequency varies depending on the status of the instrument or installation
environment. We recommend that the calibration period be determined in accordance with the
status of the instrument or installation environment. Please contact your Hioki distributor to have
your instrument periodically calibrated.
15.1 Troubleshooting
If damage is suspected, check the “Troubleshooting” section before contacting your authorized
Hioki distributor or reseller.
Before having the instrument repaired
Symptom Check Item, or Cause Remedy and Reference
Unable to write data to the
SD memory card.
Unable to manipulate
folders and les, or format
the card.
The write-protected lock of
the SD memory card is in
an intermediate position.
Check the position of the write-protect lock and
disengage it.
The connector of the SD memory card is used to judge
whether the card is write-protected.
If the write-protected lock is in an intermediate position,
the determination of whether the card is write-protected
will depend on the connector.
For example, even if the instrument determines that the
card is not write-protected and allows data to be written
to it, a computer may determine that it is write-protected,
preventing data from being written to it.
“2.4 Inserting the SD Memory Card” (p. 42)
The display does not
appear when you turn the
power on.
If powering the instrument
with the AC adapter
• Are the power cord and
AC adapter connected
properly?
Verify that the power cord and AC adapter are connected
properly.
“2.5 Power Supply” (p. 43)
If powering the instrument
with the battery
• Has the Z1003 Battery
Pack been properly
installed?
• Has the Battery Pack
been charged properly?
Verify that the battery pack has been charged and
installed.
“Installing the battery pack” (p. 38)
Keys do not work. • Has the key lock been
activated?
Press and hold the ESC key for at least 3 s to cancel the
key lock.
15 Maintenance and Service
15