195
Cleaning
Part Life Notes
Model Z1003 Battery Pack Approx. 1 year
or approx. 500
recharge/discharge
cycles whichever
comes rst
Requires periodic replacement.
15.2 Cleaning
• To clean the instrument, wipe it gently with a soft cloth moistened with water or mild detergent.
• Wipe the LCD gently with a soft, dry cloth.
15.3 Error Indication
<System Error>
Code Error indication Cause Corrective action/more information
SY01 The internal
programming of the
PQ3100 is corrupted and
the instrument must be
repaired.
The internal programming
of the instrument is corrupt.
The instrument must be repaired.
Contact your authorized Hioki distributor or
reseller.
SY02 The SDRAM of the
PQ3100 is corrupted and
the instrument must be
repaired.
The memory of the
instrument is corrupt.
SY03 The adjustment values of
the PQ3100 are corrupted
and the instrument must
be repaired.
The adjustment values of
the instrument are corrupt.
SY04 The display memory of
the PQ3100 is corrupted
and the instrument must
be repaired.
The display memory of the
instrument is corrupt.
SY05 BACKUP ERROR.
The PQ3100 must be
returned to default
factory condition.
Initialize?
YES: ENTER key
Backed-up system
variables are incorrect or
contradictory.
Initialize and recongure the settings.
If you experience backup errors frequently,
the backup battery may have deteriorated.
The instrument needs to be repaired.
Contact your authorized Hioki distributor or
reseller.
15
Maintenance and Service