162
Input Specications/Output Specications/Measurement Specications
12. Frequency 10 s measurement specications Freq10s
Measurement method Reciprocal method
Frequency calculated from the reciprocal of the accumulated time for 10 s on U1
(reference channel)
Displayed items Frequency
Measurement range 70.000 Hz
Measurement accuracy ±0.010 Hz or less
13. Frequency 200 ms measurement specications Freq
Measurement method Reciprocal method
Value calculated from the reciprocal of the accumulated time for 20 ms on U1
Displayed items Frequency
Measurement range 70.000 Hz
Measurement accuracy ±0.020 Hz or less
Event threshold Specied in deviation, 0.1 Hz to 9.9 Hz, 0.1 Hz increments
Event IN Starting time of an approximately 200-ms aggregation period to which a value belongs
when it exceeds the positive threshold value or falls below the negative threshold value
Event OUT Starting time of an approximately 200-ms aggregation period to which a frequency
value belongs when it falls below the value calculated by subtracting 0.1 Hz from
the positive threshold value or exceeds the value calculated by adding 0.1 Hz to the
negative threshold value
Equivalent to a frequency hysteresis of 0.1 Hz
Multiple-phase system
treatment
None
Saved waveforms Event waveforms
14. Voltage waveform peak measurement specications Upk
Measurement method Maximum point and minimum point of the sampled data during the 200-ms aggregation
period
Displayed items Positive waveform peak value
Negative waveform peak value
Measurement range ±2200.0 V pk
Measurement accuracy With a voltage of between 10% to 150% of the declared voltage input: 5% of declared
input voltage value
With a voltage other than above input: 2% f.s.
15. Current waveform peak measurement specications Ipk
Measurement method Maximum point and minimum point of the sampled data during the 200-ms aggregation
period
Displayed items Positive waveform peak value
Negative waveform peak value
Measurement range Ranges added the crest factor to each of the current ranges
Measurement accuracy With a current of 50% f.s. or more input: 5% rdg. + (current sensor accuracy)
With a current of less than 50% f.s. input: 2% f.s.+ (current sensor accuracy)