Specications of Input, Output, and Measurement
10.2 Specications of Input, Output, and
Measurement
Basic specications
(1) Specications common to voltage, current, power measurement
Number of input
modules
Up to 8 modules (co-existence of dierent types of input modules acceptable)
Input module type U7001 2.5MS/s Input Unit
U7005 15MS/s Input Unit
Manner of input
module installation
When dierent types of input modules coexist, install the U7005 15MS/s Input Units all
together on the Ch. 1 side.
Wiring conguration
to be measured
Single-phase 2-wire (1P2W), single-phase 3-wire (1P3W),
3-phase 3-wire (3P3W2M, 3V3A, 3P3W3M), 3-phase 4-wire (3P4W)
Wiring conguration
setting
Installed input modules can be assigned to any wiring channels. (However, only adjacent
input modules can be used to the same wiring conguration).
Dierent types of input modules can coexist in the same wiring conguration.
Dierent types of current sensors cannot coexist in the same wiring conguration.
Measurement
method
Voltage/current simultaneous digital sampling with zero-crossing synchronized calculation
Sampling frequency,
sampling bit rate
U7001: 2.5 MHz, 16-bit
U7005: 15 MHz, 18-bit
Eective
measurement range
1% of range to 110% of range
Eects of conducted
radio-frequency
electromagnetic eld
For current and active power measurement, 6% of full scale or less at 3 V
(full scale means the sensor’ rated current, only when the 9272-05 is used)
Eects of radiated
radio-frequency
electromagnetic eld
For current and active power measurement, 6% of full scale or less at 10 V/m
(full scale means the sensor’ rated current, only when the 9272-05 is used)
Display range See “10.4 Detailed Specications of Measurement Parameters” (p. 203).
Measurement mode Wideband measurement mode
Data refresh interval 10 ms, 50 ms, 200 ms
The harmonic data refresh interval is specied separately.
LPF Cuto frequency fc
U7001:
500 Hz, 1 kHz, 5 kHz, 10 kHz, 50 kHz, 100 kHz, 500 kHz, o
(The 500 kHz LPF setting uses an analog circuit.)
U7005:
500 Hz, 1 kHz, 5 kHz, 10 kHz, 50 kHz, 100 kHz, 500 kHz, 2 MHz, o
(The 2 MHz LPF setting uses an analog circuit.)
Analog LPF + Digital LPF
Add
±
0.05% of reading to the accuracy except if the LPF is set to o.
When fc is set at 500 Hz or 1 kHz, add another
±
0.5% of reading.
The accuracy specications are specied for frequencies less than or equal to one tenth
the set cuto frequency.
The peak value is based on the LPF-processed values, whereas the peak-over judgment
uses not-digital-LPF processed values.