216
Specications of Equations
Wiring
Item
1P2W 1P3W 3P3W2M 3V3A 3P3W3M 3P4W
Fundamental
wave active
power
P
1(i)
of harmonic
active power
P
1(i)(i+1)
of harmonic active power
P
1(i)(i+1)(i+2)
of harmonic active power
Fundamental
wave apparent
power
Sfnd
1(i)
=
1(i)
2
1(i)
2
+
Sfnd
1(i)(i+1)
=
1(i)(i+1)
2
1(i)(i+1)
2
Sfnd
1(i)(i+1)(i+2)
=
1(i)(i+1)(i+2)
2
1(i)(i+1)(i+2)
2
+
Fundamental
wave reactive
power
Q
1(i)
×
(−1)
of harmonic
reactive power*
1
Q
1(i)(i+1)
×
(−1)
of harmonic reactive power*
1
Q
1(i)(i+1)(i+2)
×
(−1)
of harmonic reactive power*
1
Fundamental
wave power
factor*
2
λfnd
1(i)
=
si
(i)
|cos θ
1(i)
|
λfnd
1(i)(i+1)
= si
(i)(i+1)
|cos θ
1(i)(i+1)
| λfnd
1(i)(i+1)(i+2)
= si
(i)(i+1)(i+2)
|cos θ
1(i)(i+1)(i+2)
|
If equation Type 1 is selected, the polarity sign
si
is acquired based on the sign of the fundamental wave reactive
power; if equation Type 3 is selected, based on the sign of the fundamental wave active power. If equation Type 2 is
selected, results are unsigned.
*1: If equation Type 2 is selected, take the absolute value.
*2: Fundamental wave power factor is also known as displacement power factor (DPF).
Equations for motor analysis option
Measurement item Setting Equation
Voltage Analog DC
Pulse frequency Pulse Pulse frequency
Torque
Analog DC
×
(Scaling setting value)
Frequency
[(Measurement frequency) − (
fc
setting)]
×
(Rated torque value)
(
fd
setting value)
RPM
Analog DC
×
(Scaling setting value)
Pulse
60
×
(Pulse frequency)
(Pulse count setting)
si
If the direction of rotation detection is enabled in single mode, the
polarity sign
si
is acquired based on the A-phase pulse’s rising/
falling edge and the B-phase pulse’s logic level (high/low).
Motor power —
2
×
π
×
(RPM)
60
(Torque)
×
×
(Unit coecient)
The unit coecient is one if the unit of torque measurement is
newton-meter (N
m), 1/1000 if it is millinewton-meter (mN
m), and
1000 if it is kilonewton-meter (kN
m).
Slip —
2
×
60
×
(Input frequency) − |RPM|
×
(Pole number setting)
2
×
60
×
(Input frequency)
100
×
The input frequency can be selected from among fU1 to fU8 and fI1
to fI8
M
: number of samples during synchronized timing period;
s
: sample number,
A
: analog waveform