236
Dialogs
11.3 Dialogs
If damage is suspected, read “Before returning the instrument for repair” (p. 233) as well as “11.3
Dialogs” (p. 236) below to remedy the issue. If the issue cannot be resolved, please contact your
authorized Hioki distributor or reseller.
When an error is displayed on the LCD screen, repair is necessary. Contact your authorized Hioki
distributor or reseller.
Turning on the instrument while the lines to be measured are live may damage the instrument or
cause an error to be displayed. Always turn the instrument on rst and then activate power to the
lines to be measured once you have veried that no error is being displayed by the instrument.
Error messages
Dialog Remedy
The option calibration data is
corrupted.
The instrument is in need of repair. Contact your authorized Hioki
distributor or reseller.
The option conguration has
changed.
The instrument is in need of repair. Contact your authorized Hioki
distributor or reseller.
The unit calibration data is
corrupted.
The instrument is in need of repair. Contact your authorized Hioki
distributor or reseller.
The unit ID setting is incorrect.
The instrument is in need of repair. Contact your authorized Hioki
distributor or reseller.
The instrument's settings have
been initialized.
If this message appears frequently, the instrument may be in need of
repair.
Contact your authorized Hioki distributor or reseller.
The fan is broken.
The instrument is in need of repair. Contact your authorized Hioki
distributor or reseller.
Communication part of the Unit is
broken.
The instrument is in need of repair. Contact your authorized Hioki
distributor or reseller.
Warning messages
Dialog Remedy Reference page
The current sensor has
changed.
Tap the button to close the message.
Holding values... While the hold function is activated, settings
that a󰀨ect measured values cannot be
changed. If you wish to change settings,
disable the hold function.
“5.3 Hold Function”
(p. 107)
Holding peak values... While the peak hold function is activated,
settings that a󰀨ect measured values cannot
be changed. If you wish to change settings,
disable the peak hold function.
“5.4 Peak Hold Function”
(p. 109)
Integration is ongoing, the
instrument is standing by for
integration, or integration is
stopped.
If you want to reset integration while the
instrument is integrating or standing by for
integration, stop the integration, then press
the DATA RESET key.
During the integration, settings that a󰀨ect for
other measured values cannot be changed.
“3.3 Integrating Current
and Power” (p. 64)
“Integration measurement
while using the time control
function” (p. 70)
If you want to reset integration while
integration has been stopped, press the
DATA RESET key.
HIOKI PW8001A961-00

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hioki PW8001-01 and is the answer not in the manual?

Hioki PW8001-01 Specifications

General IconGeneral
BrandHioki
ModelPW8001-01
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals