95
4
Displaying Waveforms
The instrument can display voltage, current measured on all channels and motor input waveforms.
The waveform display is completely independent of power measurement.
The operation described in this chapter will not aect measured power or harmonics values.
4.1 Waveform Display Method
Displaying waveforms on the waveform (WAVE) screen
The waveform screen displays only waveforms.
Starting waveform recording
Display screen [MEAS] > [WAVE]
Settings area
Measured value display area
Setup menu area
1
Press the RUN/STOP key.
(Lit up in green)
Waveform recording begins, and the screen
display is updated. Recording will start when
the trigger is applied.
See “4.3 Recording Waveforms” (p. 102).
2
Press the RUN/STOP key again.
(Lit up in red)
Waveform recording stops, and the screen
display is no longer updated.
Waveform recording status display
The waveform recording status display provides helpful information if it takes time for the instrument to display
waveforms or if waveforms cannot be displayed.
Trigger position (p. 100)
Waveform recording status
Stop Recording has stopped.
PreTrig. The instrument is recording
pre-trigger waveforms.
Trigger The instrument is in the trigger
standby state.
Storage The instrument is recording
post-trigger waveforms.
Compress The instrument is creating
waveforms for display.
Abort The instrument is performing
processing to stop waveform
recording.
4 Displaying Waveforms
Displaying Waveforms