EasyManua.ls Logo

JUMO DICON touch - Online Parameters; Fine Adjustment of Analog Inputs

JUMO DICON touch
196 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
173
15 Online parameters
15.1 Fine adjustment
You can use customer-specific fine adjustment to correct the measured values of the analog
input. In contrast to offsetting, which is used to specify a constant correction value for the entire
characteristic line, fine adjustment can also be used to change the slope of the characteristic
line.
Setup dialog box
Parameters
Example
The temperature inside a furnace is measured with an RTD temperature probe and displayed.
Due to the temperature drift of the sensor, the true temperature (reference measurement) de-
viates from the displayed value. The amount of deviation is different at the upper and lower
measuring points, meaning that measured value offset is not suitable.
Actual zero point: 15 °C (lower displayed value)
Set zero point: 20 °C (lower reference value)
Actual end value: 70 °C (upper displayed value)
Set end value: 80 °C (upper reference value)
Parameter Selection/settings Description
Actual-Zero-point -99999 to 0 to +99999 Lower displayed value
Actual-End value -99999 to 0 to +99999 Upper displayed value
Set-Zero-point -99999 to 0 to +99999 Lower reference value
Set-End value -99999 to 0 to +99999 Upper reference value
20°C / 80°C
20°C / 80°C
15°C / 70°C
(4)
(3)
(1)
(2)
(1) Display values (2) Reference values
(3) Furnace (4) Sensor in RTD temperature probe

Table of Contents

Other manuals for JUMO DICON touch

Related product manuals