l
The counter uses the <expected> and <resolution> parameters to configure the setting for the
SENSe:FREQuency:GATE:TIME command. The table below shows the relationship between relative res-
olution and gate time in seconds, where relative resolution = <resolution> / <expected>.
Relative Resolution
Gate Time
(53220A)
Gate Time
(53230A)
≤ 1.1E-14 1000 sec 1000 sec
>1.1E-14 to ≤1.1E-13 1000 sec 100 sec
>1.1E-13 to ≤1.1E-12 100 sec 10 sec
>1.1E-12 to ≤1.1E-11 10 sec 1 sec
>1.1E-11 to ≤1.1E-10 1 sec 100 msec
>1.1E-10 to ≤1.1E-9 100 msec 10 msec
>1.1E-9 to ≤1.1E-8 10 msec 1 msec
>1.1E-8 to ≤1.1E-7 1 msec 100 usec
>1.1E-7 to ≤1.1E-6 100 usec 10 usec
>1.1E-6 100 usec 1 usec
l Use READ? or INITiate[:IMMediate] to start the measurement.
Examples
The following example configures the instrument for frequency measurements on channel 1 using default
parameters, makes two measurements, and then transfers the measurements to the instrument output
buffer.
CONF:FREQ
SAMP:COUN 2
READ?
Typical Response: +1.00230000000000E+007,+1.00170000000000E+007
The following example configures the instrument for 10-digit frequency measurements of an approx-
imately 250 MHz signal on channel 2, makes one measurement using an external trigger, and then trans-
fers the measurement to the instrument output buffer.
CONF:FREQ 250E6,1e-4,(@2)
TRIG:SOUR EXT;SLOP POS
INIT
FETCH?
Typical Response: +2.50513398300000E+008
See Also
CONFigure?
INITiate[:IMMediate]
CONFigure:FREQuency
118 Keysight 53220A/53230A Programmer's Reference