SENSe:FREQuency:BURSt:GATE:NARRow
Syntax
[SENSe:]FREQuency:BURSt:GATE:NARRow {OFF | ON}
[SENSe:]FREQuency:BURSt:GATE:NARRow?
Description
This command enables or disables narrow pulse mode when measuring the carrier frequency of a burst
(pulse frequency). Narrow pulse mode must be enabled for bursts with a duration of 10 microseconds or
less; it must be disabled for bursts with a duration longer than 20 microseconds.
Remarks
l Available only on the Keysight 53230A. Requires option 150.
l Automatic gate setup selection (see SENSe:FREQuency:BURSt:GATE:AUTO) is always used for narrow
pulse mode measurements.
l The instrument disables narrow pulse mode after a Factory Reset (*RST command) or an Instrument Pre-
set (SYSTem:PRESet command).
Return Format
The query returns the narrow pulse mode enable state: OFF or ON.
Examples
The following example makes a pulse frequency measurement of a narrow pulse on channel 3:
CONF:FREQ:BURS
FREQ:BURS:GATE:NARR ON
READ?
Typical Response: +9.30013398300000E-007
See Also
SENSe:FREQuency:BURSt:GATE:AUTO