EasyManua.ls Logo

Keysight 53230A - Input{1|2}:Filter

Keysight 53230A
518 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
INPut{1|2}:FILTer
Syntax
INPut[{1|2}]:FILTer[:LPASs][:STATe] {OFF | ON}
INPut[{1|2}]:FILTer[:LPASs][:STATe]?
Description
This command enables or disables the input channel's low-pass filter. When enabled, the input channel is
bandwidth-limited to 100 kHz.
Remarks
l INPut or INPut1 refers to channel 1; INPut2 to channel 2.
l The low pass filter cut-off frequency is 100 KHz.
l CONFigure and MEASure? do not affect the low-pass filter setting.
l The instrument disables the low-pass filter after a Factory Reset (*RST command) or an Instrument Pre-
set (SYSTem:PRESet command).
Return Format
The query returns the current low-pass filter state: 0 (OFF) or 1 (ON).
Examples
The following example makes a frequency measurement on channel 1. The input has the low-pass filter
enabled and is configured for auto-leveling with a negative slope.
INP:FILT ON
CONF:FREQ 1.2E6
INP:SLOP NEG
READ?
Typical Response: +1.21513398300000E+006
See Also
Keysight 53220A/53230A Programmer's Reference 205