CALCulate1:SCALe:UNIT:STATe
Syntax
CALCulate[1]:SCALe:UNIT:STATe {OFF | ON}
CALCulate[1]:SCALe:UNIT:STATe?
Description
This command enables or disables displaying the unit string with measurements on the front panel when
the scaling function is enabled.
Remarks
l CALCulate1:SCALe:UNIT specifies the unit string to be displayed.
l The instrument disables displaying the unit string after a Factory Reset (*RST command) or an Instru-
ment Preset (SYSTem:PRESet command).
Return Format
The query returns the current unit string display state: 0 (OFF) or 1 (ON).
Examples
The following example enables the scaling function using the equation 100x-5 and displays "PSI" with
measurements on the front panel.
CALC:SCAL:GAIN 100.0
CALC:SCAL:OFFS 5.0
CALC:SCAL:UNIT "PSI"
CALC:SCAL:UNIT:STAT ON
CALC:SCAL:STAT ON
CALC:STAT ON
See Also
CALCulate1:SCALe:STATE
CALCulate1:STATe
CALCulate1:SCALe:UNIT
CALCulate1:SCALe:UNIT:STATe