CONFigure:PWIDth:BURSt
Syntax
CONFigure:PWIDth:BURSt [<channel>]
Description
This command sets all measurement parameters and trigger parameters to default values for positive ("on
time") width measurements of a pulsed input signal.
Parameters
Name Type Range of Values Default Value
<channel> Expression (@3) (@3)
Remarks
l Available only on the Keysight 53230A. Requires option 150.
l Use READ? or INITiate[:IMMediate] to start the measurement.
Examples
The following example configures the instrument for pulse width measurements on channel 3 using
default parameters, makes two measurements, and then transfers the measurements to the instrument
output buffer.
CONF:PWID:BURS
SAMP:COUN 2
READ?
Typical Response: +5.10520000000000E-006,+5.10370000000000E-006
See Also
CONFigure?
INITiate[:IMMediate]
MEASure?
READ?