TRIGger:DELay
Syntax
TRIGger:DELay {<time> | MINimum | MAXimum | DEFault}
TRIGger:DELay? [{MINimum | MAXimum | DEFault}]
Description
This command sets the delay time in seconds between the trigger signal and enabling the gate open for
the first measurement. This may be useful in applications where the signal you want to measure is delayed
with respect to the trigger.
Parameters
Name Type Range of Values Default Value
<time> Numeric 0 to 3600 sec with
1 us resolution
MIN=0 sec
MAX=3600 sec
0.0
Remarks
l The CONFigure and MEASure? commands set the trigger delay time to 0.0 seconds.
l The instrument sets the trigger delay time to 0.0 (no delay) after a Factory Reset (*RST command) or
after an Instrument Preset (SYSTem:PRESet command).
Return Format
The query returns the delay in seconds in the form +1.00000000000000E+000.
Examples
The following example returns 10 sets of 5 frequency measurements on channel 1, using an external trig-
ger and a 1.2 msec delay to start each set of measurements.
CONF:FREQ 1.0E6,.001
SAMP:COUN 5
TRIG:COUN 10
TRIG:SOUR EXT
TRIG:DEL .0012
READ?
Typical Response: +1.00520000000000E+006, ... (50 measurements)