CALCulate1:LIMit:STATe
Syntax
CALCulate[1]:LIMit[:STATe] {OFF | ON}
CALCulate[1]:LIMit[:STATe]?
Description
This command enables or disables limit testing.
Remarks
l The CALCulate1 subsystem must also be enabled by CALCulate1:STATe for limit testing to be per-
formed.
l The CALCulate subystem and the individual calculate functions are disabled for timestamp meas-
urements.
l The instrument clears the front panel Limit annunciator and bits 11 and 12 in the Questionable Data
Register when CALCulate1:LIMit:STATe ON is sent, when the measurement function is changed, when
the INITiate:IMMediate, MEASure? or READ? commands are executed, when the CALCu-
late1:LIMit:CLEar command is executed, after a Factory Reset (*RST command) or after an Instrument
Preset (SYSTem:PRESet command).
l The instrument disables limit testing after a Factory Reset (*RST command) or an Instrument Preset
(SYSTem:PRESet command).
Return Format
The query returns the current limit state: 0 (OFF) or 1 (ON).
Examples
The following example enables limit testing of 100 frequency measurements and returns an indication
whether measurements were outside of the range of 9.9MHz to 10.1MHz. Measurements above 10.1MHz
will set bit 12 (Upper Limit Failed) of the questionable status register; measurements below 9.9MHz will
set bit 11 (Lower Limit Failed).
*CLS
STAT:PRES
CONF:FREQ 1.0E6,.001
SAMP:COUN 100
CALC:LIM:LOW 9.9E6
CALC:LIM:UPP 10.1E6
CALC:LIM:STAT ON
CALC:STAT ON
INIT