CONFigure:FREQuency:BURSt
Syntax
CONFigure:FREQuency:BURSt [<channel>]
Description
This command sets all measurement parameters and trigger parameters to default values to measure the
carrier frequency of a burst (pulse frequency).
Parameters
Name Type Range of Values Default Value
<channel> Expression (@3) (@3)
Remarks
l Available only on the Keysight 53230A. Requires option 150.
l CONFigure and MEASure disable narrow burst mode. Because of this, measurements will return
9.91E37 (Not a Number) for bursts shorter than 10 microseconds. For such bursts, you should enable
narrow burst mode by sending SENSe:FREQuency:BURSt:GATE:NARRow ON before starting meas-
urements.
l Use READ? or INITiate[:IMMediate] to start the measurement.
Examples
The following example configures the instrument for burst frequency measurements on channel 3 using
default parameters, makes two measurements, and then transfers the measurements to the instrument
output buffer.
CONF:FREQ:BURS
SAMP:COUN 2
READ?
Typical Response: +1.00230000000000E+009,+1.00170000000000E+009
See Also
CONFigure?
INITiate[:IMMediate]
MEASure?
READ?
CONFigure:FREQuency:BURSt
120 Keysight 53220A/53230A Programmer's Reference