SENSe:TOTalize:GATE:TIME
Syntax
[SENSe:]TOTalize:GATE:TIME {<time> | MINimum | MAXimum | INFinity | DEFault}
[SENSe:]TOTalize:GATE:TIME? [{MINimum | MAXimum | DEFault}]
Description
This command selects the gate time for totalize measurements.
Parameters
Name Type Range of Values Default Value
<time> Numeric 100 usec to 1000 sec with
10 usec resolution
or +9.9E+37 (INFinity) (Keysight 53220A)
1 usec to 1000 sec with
1 usec resolution
or +9.9E+37 (INFinity) (Keysight 53230A)
MIN=100 usec (Keysight 53220A)
MIN = 1 usec (Keysight 53230A)
MAX=1000 sec
.1 sec
Remarks
l This setting is used only SENSe:TOTalize:GATE:SOURce when is set to TIME.
l When INFinity or +9.9E+37 is selected, the gate remains open until an ABORt or *RST command or a
device clear is received.
l The instrument sets the gate time to .1 sec after a Factory Reset (*RST command) or an Instrument Pre-
set (SYSTem:PRESet command).
Return Format
The query returns the gate time in the form +1.00000000000000E-001. For a infinite gate
time, the query returns 9.9E+37.
Examples
The following example makes an internally-gated totalize measurement on channel 1 with a 1 second
gate time:
FUNC "TOT"
TOT:GATE:SOUR TIME
SENSe:TOTalize:GATE:TIME
374 Keysight 53220A/53230A Programmer's Reference