CALCulate2:TRANsform:HISTogram:RANGe:AUTO:COUNt
Syntax
CALCulate2:TRANsform:HISTogram:RANGe:AUTO:COUNt {<value> | MINimum | MAXimum | DEFault}
CALCulate2:TRANsform:HISTogram:RANGe:AUTO:COUNt? [{MINimum | MAXimum | DEFault}]
Description
This command sets the number of measurements used for automatic selection of the lower and upper
range values for the histogram.
Parameters
Name Type Range of Values Default Value
<value> Numeric 10 to 1000
MIN = 10
MAX = 1000
100
Remarks
l The instrument sets the number of measurements used for automatic range value selection to 100 after
a Factory Reset (*RST command) or after an Instrument Preset (SYSTem:PRESet command).
Return Format
The query returns the number of measurements used for automatic range value selection in the form
+100.
Examples
The following example enables an automatically scaled, 128-bin histogram computation of 1000 fre-
quency measurements and returns the computed histogram, including the lower and upper range values,
the total measurement count, and the bin data. The automatic scaling is based on the first 100 meas-
urements.
CONF:FREQ 1.0E6,.001
SAMP:COUN 1000
CALC2:TRAN:HIST:RANG:AUTO ON
CALC2:TRAN:HIST:RANG:AUTO:COUN 100
CALC2:TRAN:HIST:POIN 128
CALC2:TRAN:HIST:STAT ON
INIT
*WAI
CALC2:TRAN:HIST:ALL?
Keysight 53220A/53230A Programmer's Reference 91
CALCulate2:TRANsform:HISTogram:RANGe:AUTO:COUNt