CALCulate1:SCALe:UNIT
Syntax
CALCulate[1]:SCALe:UNIT "<string>"
CALCulate[1]:SCALe:UNIT?
Description
This command sets the unit string that is displayed with measurements on the front panel when the scal-
ing function is enabled.
Parameters
Name Type Range of Values Default Value
"<string>" string one to four characters "" (null string)
Remarks
l CALCulate1:SCALe:UNIT:STATe controls whether the unit string is displayed.
l The instrument resets the units to the null string after a Factory Reset (*RST command) or an Instru-
ment Preset (SYSTem:PRESet command).
Return Format
The query returns the unit string in double quotes.
Examples
The following example enables the scaling function using the equation 100x-5 and displays "PSI" with
measurements on the front panel.
CALC:SCAL:GAIN 100.0
CALC:SCAL:OFFS 5.0
CALC:SCAL:UNIT "PSI"
CALC:SCAL:UNIT:STAT ON
CALC:SCAL:STAT ON
CALC:STAT ON
See Also
CALCulate1:SCALe:STATE
CALCulate1:STATe