DISPlay:DIGit:MASK
Syntax
DISPlay[:WINDow]:DIGit:MASK {<digits> | MINimum | MAXimum | DEFault}
DISPlay[:WINDow]:DIGit:MASK? [{MINimum | MAXimum | DEFault}]
Description
This command sets the number of significant digits displayed for measurements on the front panel. This
command does not affect the measurement resolution; it only selects the number of digits shown on the
front panel.
Parameters
Name Type Range of Values Default Value
<digits> Numeric 3 to 15
MIN = 3
MAX = 15
15
Remarks
l Setting a digit mask value disables automatic selection of the number of significant digits displayed
(sets DISPlay:DIGit:MASK:AUTO to OFF).
l Measurement data returned by the READ?, FETCh?, and MEASure? queries is unaffected by this com-
mand.
l Totalize measurements are unaffected by this command. All digits of the totalize count are always
shown.
l The digit mask is unaffected by a Factory Reset (*RST command), or an Instrument Preset
(SYSTem:PRESet command). It is set to 15 at power-on.
Return Format
The query returns the lower limit in the form +15.
Examples
The following example sets the number of significant digits displayed on the front to 6.
DISP:DIG:MASK 6
See Also
DISPlay:DIGit:MASK:AUTO