CALCulate1:AVERage:STATe
Syntax
CALCulate[1]:AVERage[:STATe] {OFF | ON}
CALCulate[1]:AVERage[:STATe]?
Description
This command enables or disables statistics computation.
Remarks
l The CALCulate1 subsystem must also be enabled by CALCulate1:STATe for statistics computation to be
performed.
l Statistics computation is disabled for continuous totalize since this computation requires more than
one measurement to be useful.
l Statistics computation is disabled for timestamp measurements.
l The instrument clears the statistical values when CALCulate1:AVERage:STATe ON is sent, when the
measurement function is changed, when the INITiate:IMMediate, MEASure? or READ? commands are
executed, when the CALCulate1:AVERage:CLEar command is executed, after a Factory Reset (*RST
command) or after an Instrument Preset (SYSTem:PRESet command).
l The instrument disables statistics computation after a Factory Reset (*RST command) or an Instrument
Preset (SYSTem:PRESet command).
Return Format
The query returns the current statistics computation state: 0 (OFF) or 1 (ON).
Examples
The following example returns the statistics of 100 frequency measurements.
CONF:FREQ 1.0E6,.001
SAMP:COUN 100
CALC:AVER:STAT ON
CALC:STAT ON
INIT
*WAI
CALC:AVER:ALL?
Typical Response: +1.00520000E+06,+1.00512000E+05,+1.005270000E+06,+4.13500000E+01
See Also
CALCulate1:STATe
CALCulate1:AVERage:STATe
76 Keysight 53220A/53230A Programmer's Reference