EasyManua.ls Logo

Keysight 53230A - System:time

Keysight 53230A
518 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SYSTem:TIME
Syntax
SYSTem:TIME <hour>,<minute>,<second>
SYSTem:TIME?
Description
This command sets the time for the instrument's real-time clock.
Parameters
Name Type Range of Values Default Value
<hour> Numeric 0 to 23 This is a required parameter
<minute> Numeric 0 to 59 This is a required parameter
<second> Numeric 0 to 59.999 This is a required parameter
Remarks
l The real-time clock is battery-backed, and retains the date and time when power is off.
l The date and time are unaffected by a Factory Reset (*RST command) or an Instrument Preset
(SYSTem:PRESet command).
Return Format
The query returns the time in the form +13,+5,+1.000.
Examples
The following example sets the time to 1:05:01 p.m.:
SYST:TIME 13,5,1
See Also
SYSTem:DATE
SYSTem:TIME
458 Keysight 53220A/53230A Programmer's Reference