CALCulate1:SCALe:INVert
Syntax
CALCulate[1]:SCALe:INVert {OFF | ON}
CALCulate[1]:SCALe:INVert?
Description
This command enables or disables 1/x inversion of the incoming reading. This inversion is done before
the Mx-B scaling function; the scaling function thus effectively becomes M/x-B.
Remarks
l The instrument disables inversion after a Factory Reset (*RST command) or an Instrument Preset
(SYSTem:PRESet command).
Return Format
The query returns the current scale and offset state: 0 (OFF) or 1 (ON).
Examples
The following example enables the scaling function using the equation 100/x-5.
CALC:SCAL:INV ON
CALC:SCAL:GAIN 100.0
CALC:SCAL:OFFS 5.0
CALC:SCAL:STAT ON
CALC:STAT ON
See Also
CALCulate1:SCALe:FUNCtion
CALCulate1:SCALe:STATE
CALCulate1:STATe