SENSe:GATE:STOP:HOLDoff:EVENts
Syntax
[SENSe:]GATE:STOP:HOLDoff:EVENts {<count> | MINimum | MAXimum | DEFault}
[SENSe:]GATE:STOP:HOLDoff:EVENts? [{MINimum | MAXimum | DEFault}]
Description
This command selects the gate close holdoff event count when SENSe:GATE:STOP:HOLDoff:SOURce is
set to EVENts.
Parameters
Name Type Range of Values Default Value
<count> Numeric 1 to 99,999,999
MIN = 1
MAX = 99,999,99
1
Remarks
l For two-channel time interval measurements, holdoff events are counted on the stop channel. For fre-
quency ratio measurements, holdoff events are counted on the denominator channel.
l The instrument sets the gate close holdoff event count to 1 after a Factory Reset (*RST command) or an
Instrument Preset (SYSTem:PRESet command).
Return Format
The query returns the gate close holdoff event count in the form +100.
Examples
The following example makes a frequency measurement on channel 1. The gate opens .1 sec after a neg-
ative edge on the Gate In/Out BNC, and, after 5 events on the input signal, closes on the next positive
edge on the BNC.
CONF:FREQ 1.2E6
FREQ:GATE:SOUR ADV
GATE:STAR:SOUR EXT
GATE:STAR:SLOP NEG
GATE:STAR:DEL:SOUR TIME
GATE:STAR:DEL:TIME .1
GATE:STOP:SOUR EXT
GATE:STOP:SLOP POS
GATE:STOP:HOLD:SOUR EVEN
SENSe:GATE:STOP:HOLDoff:EVENts
356 Keysight 53220A/53230A Programmer's Reference