CALCulate1:LIMit:CLEar
Syntax
CALCulate[1]:LIMit:CLEar[:IMMediate]
Description
This command turns off the front panel Limit annunciator and clears bit 11 ("Lower Limit Failed") and bit
12 ("Upper Limit Failed") in the condition register of the Questionable Data Register group. The cor-
responding bits in the event register are unaffected.
Remarks
l This command does not clear measurements in the reading memory.
l The instrument turns off the front panel Limit annunciator and clears bits 11 and 12 in the condition
register of the Questionable Data Register group when CALCulate1:LIMit:STATe ON is sent, when the
measurement function is changed, when the INITiate:IMMediate, MEASure? or READ? commands are
executed, when the CALCulate1:LIMit:CLEar command is executed, after a Factory Reset (*RST com-
mand) or after an Instrument Preset (SYSTem:PRESet command).
Examples
The following example clears the limit test results.
CALC:LIM:CLE
See Also
CALCulate1:LIMit:STATe
CALCulate1:STATe
CALCulate1:LIMit:CLEar
56 Keysight 53220A/53230A Programmer's Reference