CALCulate2:TRANsform:HISTogram:POINts
Syntax
CALCulate2:TRANsform:HISTogram:POINts {<value> | MINimum | MAXimum | DEFault}
CALCulate2:TRANsform:HISTogram:POINts? [{MINimum | MAXimum | DEFault}]
Description
This command sets the number of bins between the lower and upper range values for the histogram. Two
additional bins always exist: one for measurements less than the lower range value and one for meas-
urements greater than the upper range value.
Parameters
Name Type Range of Values Default Value
<value> Numeric 10 to 1000
MIN = 10
MAX = 1000
100
Remarks
l The lower and upper range values can be specified using the CALCu-
late2:TRANsform:HISTogram:RANGe:LOWer or CALCulate2:TRANsform:HISTogram:RANGe:UPPer
commands. They are computed automatically if CALCulate2:TRANsform:HISTogram:RANGe:AUTO
ON is set.
l The instrument sets the number of bins to 100 after a Factory Reset (*RST command) or after an Instru-
ment Preset (SYSTem:PRESet command).
Return Format
The query returns the number of bins in the form +100.
Examples
The following example enables an automatically scaled, 128-bin histogram computation of 1000 fre-
quency measurements and returns the computed histogram, including the lower and upper range values,
the total measurement count, and the bin data. The automatic scaling is based on the first 100 meas-
urements.
CONF:FREQ 1.0E6,.001
SAMP:COUN 1000
CALC2:TRAN:HIST:RANG:AUTO ON
CALC2:TRAN:HIST:RANG:AUTO:COUN 100
CALC2:TRAN:HIST:POIN 128
Keysight 53220A/53230A Programmer's Reference 87