SENSe:TOTalize:DATA?
Syntax
[SENSe:]TOTalize:DATA?
Description
This command returns the current totalize count.
Remarks
l Use this command for obtaining intermediate totalize counts during a continuous or long gate time
totalize measurement.
l Unlike FETCh? or READ?, which wait for a measurement to complete before returning data, this com-
mand returns intermediate data while the measurement continues.
Return Format
The query returns the totalize count in the form +1.23450000000000E+004.
Examples
The following example configures the instrument to make a continuous totalize measurement, and
returns the current totalize count after 1 second:
CONF:TOT:CONT
INIT
<wait 1 second>
TOT:DATA?
Typical Response: +1.21510000000000E+004
See Also
FETCh?
READ?
Keysight 53220A/53230A Programmer's Reference 371