MEASure:FREQuency:RATio?
Syntax
MEASure:FREQuency:RATio? [{<expected> | MINimum | MAXimum | DEFault} [,{<resolution> | MINimum
| MAXimum | DEFault}]] [,<channel_pair>]
Description
This query sets all measurement parameters and trigger parameters to default values for frequency ratio
measurements between two channels, then immediately triggers a measurement. The results are sent dir-
ectly to the instrument output buffer.
Parameters
Name Type Range of Values Default Value
<expected> Numeric Ch1/Ch2, Ch2/Ch1:
2.8E-10 to 3.5E+9
MIN = 2.8E-10
MAX = 3.5E+9
Ch1/Ch3, Ch2/Ch3
Option 106:
1.6E-11 to 3.5
MIN = 1.6E-11
MAX = 3.5
Option 115:
6.6E-12 to 1.2
MIN = 6.6E-12
MAX = 1.2
Ch3/Ch1, Ch3/Ch2:
Option 106:
0.28 to 6.0E10
MIN = .28
MAX = 6.0E10
Option 115:
0.85 to 15.0E10
MIN = .85
MAX = 15.0E10
1.0
<resolution> Numeric 1.0E-15 * <expected> to
1.0E-5* <expected>
MIN = 1.0E-15 * <expected>
MAX = 1.0E-5 * <expected>
Value corresponding to
.1 second gate time
<channel_pair> Pair of
Expressions
(@1),(@2) | (@1),(@3) |
(@2),(@1) | (@2),(@3) |
(@3),(@1) | (@3),(@2)
(@1),(@2)