INPut{1|2}:RANGe
Syntax
INPut[{1|2}]:RANGe {<range> | MINimum | MAXimum | DEFault}
INPut[{1|2}]:RANGe? [{MINimum | MAXimum | DEFault}]
Description
This command selects the voltage range for the input channel.
Parameters
Name Type Range of Values Default Value
<range> Numeric 1:1 probe factor:
5.0, 50.0
MIN = 5.0
MAX = 50.0
10:1 probe factor:
50.0, 500.0
MIN = 50.0
MAX = 500.0
5.0
50.0
Remarks
l INPut or INPut1 refers to channel 1; INPut2 to channel 2.
l When a 1:1 probe factor is selected (INPut{1|2}:PROBe set to 1), the available ranges are 5.0 and 50.0.
When a 10:1 probe factor is selected (INPut{1|2}:PROBe set to 1), the available ranges are 50.0 and
500.0.
l CONFigure and MEASure? do not affect the input range.
l The instrument selects the 5V range after a Factory Reset (*RST command) or an Instrument Preset
(SYSTem:PRESet command).
Return Format
The query returns the input range in the form +5.00000000E+000.
Examples
The following example makes a frequency measurement on channel 1. The input is configured for the
50V range and a DC-coupled 12V threshold with a negative slope.
INP:RANG 50
INP:COUP DC
INPut{1|2}:RANGe