CONFigure:SPERiod
Syntax
CONFigure:SPERiod [<channel>]
Description
This command sets all measurement parameters and trigger parameters default values for single-period
measurements.
Parameters
Name Type Range of Values Default Value
<channel> Expression (@1) | (@2) (@1)
Remarks
l This command configures a single-period measurement. To configure an averaging measurement, use
CONFigure:PERiod.
l The input channel is configured for auto-leveling at 50% with a positive slope.
l This measurement is a specialized time interval measurement. The gate is controlled by the
SENSe:TINTerval:GATE:SOURce command.
l The gate source (see SENSe:TINTerval:GATE:SOURce) is set to IMMediate.
l Use READ? or INITiate[:IMMediate] to start the measurement.
Examples
The following example configures the instrument for single-period measurements on channel 1 using
default parameters, makes two measurements, and then transfers the measurements to the instrument
output buffer.
CONF:SPER
SAMP:COUN 2
READ?
Typical Response: +1.00230000000000E-007,+1.00170000000000E-007
See Also
CONFigure?
CONFigure:PERiod
INITiate[:IMMediate]
MEASure?
CONFigure:SPERiod
148 Keysight 53220A/53230A Programmer's Reference