DATA:LAST?
Syntax
DATA:LAST?
Description
This query returns the last measurement taken.You can execute this command at any time, even during a
series of measurements.
Remarks
l The instrument clears all measurements from memory when the measurement function is changed,
when the INITiate:IMMediate, MEASure:<function>? or READ? commands are executed, after a Factory
Reset (*RST command) or after an Instrument Preset (SYSTem:PRESet command).
Return Format
The query returns one measurement with units. If no data is available, 9.91E37 (Not a Number) is returned
with units. The format of the measurements is always ASCII (unaffected by the FORMat Subsystem com-
mands).
Examples
The following example returns the last measurement taken by the meter.
DATA:LAST?
Typical Response: +1.21513398300000E-006 S