INPut{1|2}:LEVel{1|2}
Syntax
INPut[{1|2}]:LEVel[{1|2}][:ABSolute] {<volts> | MINimum | MAXimum | DEFault}
INPut[{1|2}]:LEVel[{1|2}][:ABSolute]? [{MINimum | MAXimum | DEFault}]
Description
This command sets the input threshold voltage for measurements. Setting the input threshold voltage dis-
ables auto-leveling (sets INPut{1|2}:LEVel:AUTO to OFF).
If auto-leveling is enabled (INPut{1|2}:LEVel:AUTO ON set), this value will be updated each time a leveling
operation occurs, and will reflect the current threshold voltage.
LEVel2 may only be queried when the measurement is rise/fall time or single-channel time interval; oth-
erwise, querying LEVel2 will also return 9.91E+37.
Parameters
Name Type Range of Values Default Value
<value> Numeric 5V range:
-5.125 V to +5.125 V
with 2.5 mV resolution
MIN = -5.125 V
MAX = +5.125 V
50V range:
-51.25 V to +51.25 V
with 25 mV resolution
MIN = -51.25 V
MAX = +51.25 V
500V range:
-512.5 V to +512.5 V
with 250 mV resolution
MIN = -512.5 V
MAX = +512.5 V
0.0
Remarks
l INPut or INPut1 refers to channel 1; INPut2 to channel 2.
l When a 1:1 probe factor is selected (INPut{1|2}:PROBe set to 1), the available ranges are 5.0 and 50.0.
When a 10:1 probe factor is selected (INPut{1|2}:PROBe set to 1), the available ranges are 50.0 and
500.0.
Keysight 53220A/53230A Programmer's Reference 213
INPut{1|2}:LEVel{1|2}