CONFigure:PDUTycycle
Syntax
CONFigure:PDUTycycle [{<reference> | MINimum | MAXimum | DEFault}] [,<channel>]
Description
This command sets all measurement parameters and trigger parameters to default values for positive
duty cycle measurements.
Parameters
Name Type Range of Values Default Value
<reference> Numeric 10 to 90 PCT (percent)
-5.125 to +5.125 V (5.125V range)
-51.25 to +51.25 V (51.25V range)
MIN = 10 PCT
MAX = 90 PCT
50 PCT
<channel> Expression (@1) | (@2) (@1)
Remarks
l The <reference> parameter specifies the input signal reference level, either in terms of percent of peak-
to-peak voltage, or in absolute voltage. To specify the level in percent, use a numeric value with no suf-
fix or with the PCT suffix; for example, 30 or 30 PCT. To specify the level in absolute voltage, use a
numeric value with the V or MV (millivolt) suffix: 100 MV or .1V.
l When the <reference> is omitted or specified in percent, auto-leveling is enabled. When specified in
absolute voltage, auto-leveling is disabled.
l Use READ? or INITiate[:IMMediate] to start the measurement.
l Measurements are returned as a fraction (not as percent).
Examples
The following example configures the instrument for pulsed duty cycle measurements on channel 1 using
default parameters, makes two measurements, and then transfers the measurements to the instrument
output buffer.
CONF:PDUT
SAMP:COUN 2
READ?
Typical Response: +5.10520000000000E-001,+5.10370000000000E-001
CONFigure:PDUTycycle
136 Keysight 53220A/53230A Programmer's Reference