CALibration:ALL?
Syntax
CALibration[:ALL]?
Description
This query performs a calibration of the instrument using the specified calibration value (CALibra-
tion:VALue command). Before you can calibrate the instrument, you must unsecure it by entering the cor-
rect security code.
For a more detailed discussion of the calibration procedures, see the Keysight
53210A/53220A/53230A 350 MHz Counter Service Guide. Please refer to the Ser-
vice Guide before attempting to calibrate the instrument. Improper use of the
CALibration commands can adversely affect the accuracy and reliability of the instru-
ment.
Remarks
l If a calibration fails, "+1" is returned and an error is stored in the error queue. For a complete listing of
the error messages related to calibration failures, see SCPI Error Messages.
l This query increments the calibration count on the counter (see CALibration:COUNt? command).
Return Format
The query returns +0 (calibration passed) or +1 (calibration failed).
Examples
The following example performs a calibration and returns a pass/fail indication.
CAL?
Typical Response: +0
See Also
CALibration:SECurity:STATe
CALibration:VALue
CALibration:ALL?
100 Keysight 53220A/53230A Programmer's Reference